WEKO3
アイテム
{"_buckets": {"deposit": "cc055605-92dd-4c57-8b1c-89eee9fad567"}, "_deposit": {"created_by": 3, "id": "1000", "owners": [3], "pid": {"revision_id": 0, "type": "depid", "value": "1000"}, "status": "published"}, "_oai": {"id": "oai:mdu.repo.nii.ac.jp:00001000", "sets": ["88"]}, "author_link": ["0"], "item_2_alternative_title_1": {"attribute_name": "その他(別言語等)のタイトル", "attribute_value_mlt": [{"subitem_alternative_title": "Application of Ion-plating Technique for Orthodontic Stainless Steel Arch Wire"}]}, "item_2_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "1992-12-31", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "3", "bibliographicPageEnd": "276", "bibliographicPageStart": "271", "bibliographicVolumeNumber": "18", "bibliographic_titles": [{"bibliographic_title": "松本歯学"}]}]}, "item_2_description_19": {"attribute_name": "フォーマット", "attribute_value_mlt": [{"subitem_description": "application/pdf", "subitem_description_type": "Other"}]}, "item_2_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "The purpose of this study was the development of esthetic orthodontic wire. In this study, the authors performed the technique of applying Ti-N Ion-plating to orthodontic stainless steel arch wire. The authors also compared Ion-plated stainless steel wire with the original one by performing a tensile test, and three pointed bending test, as well as test of surface roughness and Vicker\u0027s hardness. The conclusion is as follows: 1: A layer of Ion-plate (2μm thickness) was observed in the original stainless steel wire, but the coating layer was not observed an omission or removal with optical or electron microscopes. 2: There was no statistical difference between Ion-plated stainless steel wire and the original one in terms of tensile strength, yield strength, elongation, bending stress, proof stress and surface roughness. 3: A slight difference between Ion-plated stainless steel wire and the original one was observed in the Vicker\u0027s hardness test. These results suggest that Ti-N coating produced by means of the Ion-plating technique is effective for clinical use of orthodontic arch wire.", "subitem_description_type": "Abstract"}]}, "item_2_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "松本歯科大学学会"}]}, "item_2_source_id_11": {"attribute_name": "書誌レコードID", "attribute_value_mlt": [{"subitem_source_identifier": "AN00232590", "subitem_source_identifier_type": "NCID"}]}, "item_2_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0385-1613", "subitem_source_identifier_type": "ISSN"}]}, "item_2_version_type_20": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "白井, 竹郎"}, {"creatorName": "小幡, 明彦"}, {"creatorName": "宮崎, 顕道"}, {"creatorName": "大山, 尚毅"}, {"creatorName": "丹羽, 健"}, {"creatorName": "戸苅, 惇毅"}, {"creatorName": "出口, 敏雄"}, {"creatorName": "山岸, 利夫"}, {"creatorName": "伊藤, 充雄"}], "nameIdentifiers": [{"nameIdentifier": "0", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2012-05-24"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "matsumoto_shigaku_18-03-06.pdf", "filesize": [{"value": "860.0 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 860000.0, "url": {"label": "matsumoto_shigaku_18-03-06.pdf", "url": "https://mdu.repo.nii.ac.jp/record/1000/files/matsumoto_shigaku_18-03-06.pdf"}, "version_id": "dcf16c8f-6405-4f13-9362-aae9b2f90551"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "Ti-N Ion-plating", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}, {"subitem_subject": "orthodontic stainless steel wire", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}, {"subitem_subject": "esthetic", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "矯正用材料へのイオン・プレーティングの応用(その1) : アーチワイヤーへの応用", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "矯正用材料へのイオン・プレーティングの応用(その1) : アーチワイヤーへの応用"}]}, "item_type_id": "2", "owner": "3", "path": ["88"], "permalink_uri": "https://mdu.repo.nii.ac.jp/records/1000", "pubdate": {"attribute_name": "公開日", "attribute_value": "2012-05-24"}, "publish_date": "2012-05-24", "publish_status": "0", "recid": "1000", "relation": {}, "relation_version_is_last": true, "title": ["矯正用材料へのイオン・プレーティングの応用(その1) : アーチワイヤーへの応用"], "weko_shared_id": -1}
矯正用材料へのイオン・プレーティングの応用(その1) : アーチワイヤーへの応用
https://mdu.repo.nii.ac.jp/records/1000
https://mdu.repo.nii.ac.jp/records/1000062f7fb9-03b8-4614-bb8e-c4f79a56036c
名前 / ファイル | ライセンス | アクション |
---|---|---|
matsumoto_shigaku_18-03-06.pdf (860.0 kB)
|
|
Item type | Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2012-05-24 | |||||
タイトル | ||||||
タイトル | 矯正用材料へのイオン・プレーティングの応用(その1) : アーチワイヤーへの応用 | |||||
言語 | ||||||
言語 | jpn | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | Ti-N Ion-plating | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | orthodontic stainless steel wire | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | esthetic | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
その他(別言語等)のタイトル | ||||||
その他のタイトル | Application of Ion-plating Technique for Orthodontic Stainless Steel Arch Wire | |||||
著者 |
白井, 竹郎
× 白井, 竹郎 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The purpose of this study was the development of esthetic orthodontic wire. In this study, the authors performed the technique of applying Ti-N Ion-plating to orthodontic stainless steel arch wire. The authors also compared Ion-plated stainless steel wire with the original one by performing a tensile test, and three pointed bending test, as well as test of surface roughness and Vicker's hardness. The conclusion is as follows: 1: A layer of Ion-plate (2μm thickness) was observed in the original stainless steel wire, but the coating layer was not observed an omission or removal with optical or electron microscopes. 2: There was no statistical difference between Ion-plated stainless steel wire and the original one in terms of tensile strength, yield strength, elongation, bending stress, proof stress and surface roughness. 3: A slight difference between Ion-plated stainless steel wire and the original one was observed in the Vicker's hardness test. These results suggest that Ti-N coating produced by means of the Ion-plating technique is effective for clinical use of orthodontic arch wire. | |||||
書誌情報 |
松本歯学 巻 18, 号 3, p. 271-276, 発行日 1992-12-31 |
|||||
出版者 | ||||||
出版者 | 松本歯科大学学会 | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0385-1613 | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN00232590 | |||||
フォーマット | ||||||
内容記述タイプ | Other | |||||
内容記述 | application/pdf | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |