@article{oai:mdu.repo.nii.ac.jp:00001284, author = {舛田, 篤之 and 村上, 弘 and 神谷, 光男 and 大和, 篤弘 and 橋本, 京一}, issue = {3}, journal = {松本歯学}, month = {Dec}, note = {application/pdf, In order to accumulate the fundamental data for discussing the thermal distribution of shape memory blade implant by high frequency heating, three points on the implant body were measured by micro thermocouples. The results were as follows: 1. When the heating condition was 2.5A 30 sec, the temperature was too low for the shape memory effect, but when the heating condition was 3A 30 sec and 3.5A 30 sec, the temperature on the neck part of the implant rose too high. 2. When the temperature was 50.7℃ (mean) on the neck part and 40.0℃ (mean) on the end of the foot part of the implant with a heating condition of 2.5A 60 sec, the shape memory effect was obtained. 3. When we used high power-short time heating, only the surface of the neck part reached a high temperature. Because the thermal conductivity of NT alloy (Nitinol) was low, however, the end of foot part didn't heat. Satisfactory results were obtained when low power-long time heating was used.}, pages = {288--292}, title = {高周波加温による形状記憶インプラントの温度分布について}, volume = {12}, year = {1986} }