@article{oai:mdu.repo.nii.ac.jp:00000915, author = {池谷, 虎彦 and 鈴木, 寿典 and 中阪, 雅昭 and 土井, 久栄 and 関澤, 俊郎 and 行木, 貴宏 and 笠原, 悦男 and 安田, 英一}, issue = {2}, journal = {松本歯学}, month = {Aug}, note = {application/pdf, Today, it is necessary for dentists to wear laxtex gloves for cross infection control in general dental practices. However, some problems have arisen because of the use of latex gloves. One of these such problems has occurred during electric pulp tests. When electric pulp testing occurs, the circuit is completed by the dentists placing one hand on the electrode handle and one or two fingers in contact with the patient's cheek or lip. The circuit was severed by the wearing of latex gloves. To solve this problem, the patient grasped the metal pole that was connected to the electrode handle with an electric cord. This clinical study was carried out for the purpose of knowing whether or not the measured this new electric pulp testing method differed from those of the usual method. The results obtained were as follows. 1. There were no different measured values between the new and usual methods. These values were also confirmed statistically. 2. It was recognized that the measured values had much difference among individuals.}, pages = {152--156}, title = {電気歯髄診断器無関電極の接触位置変更による測定値の変動について}, volume = {20}, year = {1994} }